This kit is for in vitro research use only, not for clinical diagnosis! Experimental principle The purified IGF-1 antibody is used to coat the microplate to make a solid phase carrier. The specimen or standard, biotinylated IGF-1 antibody, and HRP-labeled avidin are added to the microwell in sequence. After washing, it is developed with a substrate (TMB). TMB is converted into blue under the catalysis of peroxidase, and into the final yellow under the action of acid. The color depth is positively correlated with IGF-1 in the sample. The absorbance (OD value) was measured with a microplate reader at a wavelength of 450 nm, and the sample concentration was calculated. Kit composition and reagent preparation Bring your own items Collection and preservation of specimens Before starting the experiment, all reagents should be equilibrated to room temperature, and the reagents should not be dissolved in 37 directly. When preparing the reagents or samples, they must be thoroughly mixed. Try to avoid foaming when mixing. The sample content should be predicted before the experiment. If the sample concentration is too high, the sample should be diluted to make the diluted sample meet the detection range of the kit, and then multiplied by the corresponding dilution factor when calculating. Note: Specificity This kit can detect recombinant or natural mouse IGF-1 at the same time, and does not cross-react with other related proteins. Calculate the OD value of each standard and sample after deducting the OD value of the blank hole (seven-point diagram). If multiple holes are set, the average value should be used for calculation. Taking the concentration of the standard product as the ordinate (logarithmic coordinate) and the OD value as the abscissa (logarithmic coordinate), a standard curve is drawn on the logarithmic coordinate paper. It is recommended to use professional curve making software for analysis, such as curve expert 1.3, according to the OD value of the sample, find the corresponding concentration from the standard curve and multiply it by the dilution factor; or use the concentration and OD value of the standard to calculate the regression equation of the standard curve, Substitute the OD value of the sample into the equation, calculate the sample concentration, and then multiply it by the dilution factor to obtain the actual concentration of the sample. Detection range: 0.039 ng / ml-2.5 ng / ml, please use the following concentration values ​​for drawing standard curve: 2.5 ng / ml, 1.25 ng / ml, 0.625 ng / ml, 0.312 ng / ml, 0.156 ng / ml, 0.078 ng / ml, 0.039 ng / ml. Minimum detection limit: 0.0195 ng / mL Explanation Linte remover,Best Seller Lint Removers,Best Seller Lint Removers,Clothing Lint Removers Zhejiang Hisun Electrical Appliance Co.,Ltd , https://www.cn-hisun.com
Intended application
Quantitative determination of insulin-like growth factor 1 (IGF-1) in mouse serum, plasma or other related biological fluids by ELISA.
1. Microplate: one piece (96 wells)
2. Standard product (lyophilized product): 2 bottles, please prepare within 15 minutes before use. Each bottle is diluted to 1ml with sample diluent. After capping, it is allowed to stand at room temperature for about 10 minutes. At the same time, it is inverted / rubbed repeatedly to help dissolution. Its concentration is 10 ng / ml. Multiple dilutions (Note: Do not directly perform multiple dilutions in the plate), formulated into 2.5 ng / ml, 1.25 ng / ml, 0.625 ng / ml, 0.312 ng / ml, 0.156 ng / ml, 0.078 ng / ml, 0.039 ng / ml, the sample dilution is directly used as blank well 0 ng / ml. If preparing 1.25 ng / ml standard: take 0.5ml (not less than 0.5ml) of the above standard at 2.5 ng / ml and add it to the Eppendorf tube containing 0.5ml of sample diluent, mix well.
3. Sample diluent: 1 × 20ml.
4. Test dilution A: 1 × 10ml.
5. Test diluent B: 1 × 10ml.
6. Detection solution A: 1 × 120 per bottle (1: 100). Before use, dilute with Test Diluent A 1: 100 (for example: 10 Test Solution A / 990 Test Diluent A), mix well, and prepare according to the pre-calculated total amount required for each experiment (100 / Hole), more 0.1-0.2ml should be prepared during actual preparation.
7. Detection solution B: 1 × 120 per bottle (1: 100). Dilute 1: 100 with test dilution B immediately before use. The dilution method is the same as that of Test Solution A.
8. Substrate solution: 1 × 10ml / bottle.
9. Concentrated washing liquid: 1 × 30ml / bottle, each bottle is diluted 25 times with distilled water.
10. Stop solution: 1 × 10ml / bottle (2 mol / L H2SO4).
11. Lamination: 5 sheets
12. Instruction manual: 1 copy
1. Microplate reader (It is recommended to preheat the instrument before use)
2. Micro dosing device and pipette tip, EP tube
3. Distilled or deionized water, filter paper
1. Serum: Whole blood samples should be left at room temperature for 2 hours or 4 overnight. After centrifugation at 1000 g for 20 minutes, take the supernatant for detection, or store the supernatant at -20 or -80, but avoid repeated freezing and thawing. .
2. Plasma: EDTA or heparin can be used as an anticoagulant. Centrifuge the sample at 1000 g for 15 minutes within 30 minutes after collection. The supernatant can be taken for detection, or the supernatant should be stored at -20 or -80, but avoid repeated Freeze and thaw.
3. Other biological specimens: centrifuge at 1000 g for 20 minutes, take the supernatant for testing, or store the supernatant at -20 or -80, but avoid repeated freezing and thawing.
Note: The above specimens should be sealed and stored, 4 should be stored for less than 1 week, -20 should not exceed 1 month, -80 should not exceed 2 months; specimen hemolysis will affect the final test results, so hemolysis specimens should not be tested .
Sample processing: serum or plasma specimens are recommended to be diluted 10-100 times, for example: diluted 10 times, 100uL serum or plasma is added to 900uL PBS. Specimens were diluted with 0.1 M PBS (PH = 7.0-7.2). It is recommended that each laboratory conduct pre-experiment before operation to establish the optimal dilution factor.
1. Add sample: set blank hole, standard hole and sample hole to be tested respectively. Add 100% of the sample diluent to the blank well, and 100% of the standard or the sample to be tested in the remaining well. Be careful not to have air bubbles. Add the sample to the bottom of the enzyme plate when adding the sample. Try not to touch the wall of the well. Gently shake to mix the enzyme. Cover or cover the target plate and incubate at 37 for 2 hours. To ensure the validity of the experimental results, please use a new standard solution for each experiment.
2. Discard the liquid and dry it without washing. Add 100 working solutions of detection solution A (prepared before use) to each well, add the microplate to the membrane, and incubate at 37 for 1 hour.
3. Discard the liquid in the hole, spin dry, wash the plate 3 times, soak for 1-2 minutes each time, about 400 per hole, spin dry (you can also pat dry the liquid in the hole).
4. Add 100% of the working solution of the test solution B (prepared before use) to each well, add a coating, and incubate at 37 for 1 hour.
5. Discard the liquid in the hole, spin dry, wash the plate 5 times, the method is the same as step 3.
6. Add substrate solution 90 per well, enzyme label plate and film 37 to avoid light and color development (reaction time is controlled at 15-30 minutes, when the first 3-4 wells of the standard well have a clear blue gradient, the last 3 -When the gradient of 4 wells is not obvious, it can be terminated).
7. Add 50% stop solution to each well to stop the reaction. At this time, the blue color turns to yellow. The order of adding the stop solution should be the same as that of the substrate solution.
8. Immediately measure the optical density (OD value) of each well with a microplate reader at a wavelength of 450 nm.
1. Preparation of reagents: All reagents should be equilibrated to room temperature before use. Please save the reagents according to the instructions immediately after use. Please use disposable tips during the experiment to avoid cross contamination.
2. Sample addition: when adding samples or adding reagents, if the time interval between the first well and the last well is too large, it will result in different "pre-incubation" time, which will obviously affect the measured value Accuracy and repeatability. The time of one sample addition (including standard and all samples) is best controlled within 10 minutes. It is recommended to set up multiple holes for experiments.
3. Incubation: In order to prevent the sample from evaporating, place the enzyme-labeled plate with a cover or membrane in the wet box during the experiment to avoid liquid evaporation; the next step should be performed as soon as possible after washing the plate, and should be avoided at any time The microplate is in a dry state; at the same time, the given incubation time and temperature should be strictly observed.
4. Washing: The washing liquid remaining in the reaction well during the washing process should be patted dry on the filter paper. Do not put the filter paper directly into the reaction well to absorb water. At the same time, the remaining liquid and fingerprints on the bottom of the plate should be eliminated to avoid affecting the final enzyme label. Instrument reading.
5. Preparation of reagents: before using, please shake the hands a few times or centrifuge for a while to prevent the liquid on the tube wall or bottle cap from depositing to the bottom of the tube. Standard products, working solution A, and working solution B should be prepared according to the required amount, and prepared with the corresponding diluent, not to be confused. Please accurately prepare the standard and working solution, and try not to prepare in small amounts (such as when drawing the test solution A, it should not be less than 10 at a time) to avoid concentration errors due to inaccurate dilution; do not reuse the diluted standard and test Solution A working fluid, detection solution B working fluid.
6. Control of reaction time: Please observe the color change of the reaction well regularly after adding the substrate (for example, every 10 minutes). If the color is dark, please add the stop solution in advance to stop the reaction to avoid the reaction being too strong and affecting the microplate reader. Optical density reading.
7. Substrate: Please keep the substrate away from light, and avoid direct exposure to strong light during storage and incubation.
Washing method
1. Manual plate washing method: Inject at least 0.4ml of the recommended washing buffer into the wells, soak for 1-2 minutes, suck (do not touch the wall) or shake off the liquid in the enzyme plate, and put a few layers on the experimental table Absorbent paper, with the microtiter plate down and vigorously pat several times; repeat this process several times as needed.
2. Automatic plate washing: If there is an automatic plate washing machine, it should be used in the formal experiment process after being used proficiently.
1. Due to the existing conditions and the level of science and technology, it is not possible to conduct a comprehensive identification and analysis of all raw materials provided by all suppliers,
This product may have certain quality and technical risks.
2. The final experimental result is closely related to the effectiveness of the reagents, the relevant operations of the experimenter and the experimental environment at that time. Please be sure to prepare sufficient specimen backups.
3. Only all USCNLIFETM reagents can be used to ensure the detection effect, because all reagents are related and cannot be mixed with other manufacturers' products. Only by strictly following the experimental instructions of the USCNLIFETM reagent will the best test results be obtained. If the result of the experimenter's mistakes or improper storage of the kit results in poor results, it is not a product quality problem.
4. Avoid exposing the reagent to strong light during storage and incubation. All reagent bottle caps must be tightly closed to prevent evaporation and microbial contamination, because the interference of proteolytic enzymes will lead to erroneous results.
5. Storage of the kit: Please save the standard, detection solution A and detection solution B at -20 as soon as possible after receiving the kit. The short-term storage of the remaining reagents should be placed at 4, and the long-term storage at -20. After opening, the microplate should be sealed with desiccant and stored at -20 to avoid moisture.
6. Salt will be precipitated from the concentrated washing liquid, which can be heated and dissolved in the water bath when diluted.
7. There may be a little water-like substance in the well of the enzyme-linked plate just opened. This is a normal phenomenon and will not have any impact on the experimental results.
8. Validity: 6 months.
9. These operating instructions apply to the 48T kit, but all reagents in the 48T kit are halved.